Dust removal in temperature and humidity Test Chamber (1/2)

The condenser should be maintained regularly and monthly, and dust attached to the condenser heat dissipation mesh should be removed by vacuum cleaner or dust by high pressure air spray.

The repalcement the wet ball test cloth in temperature and humidity test chamber.
-When its surface is not clean or hard, or after the temperature control is finished, the test cloth must be replaced before continuing to do the temperature and humidity control. The test cloth should be replaced once every three months. When changing, clean cloth should be used to wipe the temperature measuring unit [Sen-sor]. When replacing the new test cloth, it should be cleaned first.

The inspection and adjustment of the wet ball water level in the chamber
-The water level in the tank should not be too high, so that the water overflow water tube or too low makes the wet ball test cloth water absorption abnormal, and the accuracy of the wet ball water level can be kept about 60% full. 