Call Us Now! +86-551-63853683 Sales@climatechambers.com
You are here:Home >> Support >> Latest News

News

News

Contact Us

Symor Instrument Equipment Co.,Ltd
Email: Sales@climatechambers.com
Tel: +86-551-63853683, 68663630
Shanghai office 
No.3215,Huhang Highway,Fengxian District, Shanghai, 231417,China
Hefei office
No.105 Dayang Industrial Park
Luyang District, Hefei, 230041, China

Latest News

How to maintain constant temperature humidity chamber?

The following suggestions are for reference.

1. cleaning the condenser regularly (every 3 months): for the compressor using air-cooled cooling, the condenser should be regularly overhauled, and the condenser should be decontaminated to ensure its good ventilation and heat transfer performance; for the compressor using water-cooled cooling, in addition to ensuring its influent pressure, influent temperature, it must also ensure the corresponding flow rate, and regularly clean and scale the condenser to obtain its continuous heat transfer performance.

2. regular (every 3 months) cleaning evaporation (dehumidifier): because the cleanliness grade of the sample is different, under the action of forced wind circulation, evaporation (dehumidification) will condense a lot of dust and other small particle objects, should be cleaned regularly.
                     

3. because the price of the temperature (humidity) chamber is generally high, we suggest that it be placed in a relatively benign temperature environment. Our empirical temperature is 5℃~30℃. For laboratories that do not have this condition, Appropriate air conditioners (air cooling) should be provided.

4. adhere to professional management. Conditional units, should send personnel to the supplier factory training and study, in order to obtain more professional maintenance, maintenance experience and ability.

5. circulating air blade, condenser fan cleaning and balance: similar to the cleaning evaporator, because the working environment of the test box is different, the circulating air blade, condenser fan will condense a lot of dust and other small particle objects, should be cleaned regularly.

6. waterway, humidifier cleaning: if the waterway is not smooth, humidifier scaling is easy to lead to humidifier dry burning, may damage humidifier, so must regularly clean the waterway, humidifier.

7.After each test,set the temperature near the ambient temperature, work for about 30 minutes, then cut off the power supply, and clean the inner wall of the studio.

8. If the temperature humidity test chamber is long-term not used, the product should be regularly electrified every half a month, the electrification time is not less than 1 hour.

In a summary, there is detailed maintenance procedures and attentions about temperature humidity cabinet in our operation manual, we will guide customers about maintenance issues, customer can take easy to purchase.

Copyright © 2001-2021 Symor Instrument Equipment Co., Ltd. All Rights Reserved